- Dearation, preheating and precise dosing of dissolution media
- Volumetric dispensing with an accuracy better than +/- 1% as per USP and Ph. Eur. requirements
- Mobile unit allows media filling in the test vessel within the dissolution bath
- Multifunction single hose for dispensing, aspiration, cleaning, rinsing and drainage
- Ergonomically designed nozzle with embedded dispense switch for ease of operation
- User friendly calibration and validation of media volumes
- Report printout of calibration, validation and dispense cycle
- Batch wise dispensing and stores last 10 batch information of media
- Special nozzle design allows the handling of surfactants like Sodium Lauryl Sulphate (SLS)
- Programmable dearation time and temperature for effective dearation of different media
- Maintains media under vacuum at desired temperature while in idle condition
- Easy access for manual cleaning of vessel tank
- Automatic rinsing and cleaning cycle
- Virtually no dead volume
- Programmable dispensing volumes from 200 ml to 2 ltrs
- Dispensing rate 900 ml/min
Specifications
Aspiration: - Media Volume: Programmable from 0.2 litre to 21 litres
- Accuracy: +1%
- Flow rate: 900 ml/min. (During run cycle)
Dispense: - Dispense Volume: Programmable from 0.2 L to 4 L
- Accuracy: Better than +1%
- Flow rate: 900 ml/min.
Temperature: - Temperature Range: 20oC to 50oC
- Accuracy: +0.5oC
Media preparation time: 45 min. (For 21 L of media @ambient temperature 25o C)
O2 level: Normal tap water which contains 85% of dissolved O2 before degassing will contain 40% of dissolved O2 after degassing (Measured by DO2 meter)
Hose: Single (For aspiration, dispense and drain)
Media tank: S.S. 316, Teflon coated
Vacuum pump: 760 mmHg
Piston pump: Ceramic metering piston pump
Accuracy: Better than +1%
Power: 1200 watts
Power supply: 230V AC/50 Hz. +10%
Ambient conditions: - Humidity: 70% max.
- Storage temperature: 10o C to 50o C
- Operating temperature: 10o C to 30o C
Dimensions: (ǿ 554 x 920 H) mm
Weight: - Net 100 kgs
- Gross 132 kgs